Your browser doesn't support javascript.
loading
Research of radiation dose rate around patients underwent renal dynamic imaging of SPECT with 99Tcm-DTPA / 中国医学装备
China Medical Equipment ; (12): 23-26, 2017.
Article Dans Chinois | WPRIM | ID: wpr-613201
ABSTRACT

Objective:

To study the radiation dose rate around patients underwent renal dynamic imaging of SPECT with 99Tcm-DTPA, so as to provide the basis for ensuring the radiation safety of the surrounding environment and the general public.

Methods:

109 patients who underwent renal dynamic imaging were enrolled in this study, and the radiation dose rates of different time and different distance after 99Tcm-DTPA was injected patients were measured. And then, the changes of dose rate following time and distance were analyzed.

Results:

At the 0.5h, 1h, 2h, 3h, 6h, and 7h after radiopharmaceuticals were injected patients underwent renal dynamic imaging, the radiation dose rates were 0.76-8.86 μSv/h, 0.68-7.27 μSv/h, 0.57-4.52 μSv/h, 0.56-3.90 μSv/h, 0.23-2.07 μSv/h and 0.21-1.05 μSv/h, respectively, between 0.5 m and 4 m away from patients. At 24h after radiopharmaceuticals were injected, all of radiation dose rates at different distances around patients were same with background values.

Conclusion:

The radiation dose rates around patients underwent renal dynamic imaging will rapidly decrease with the increasing of time and distance. At 24h after radiopharmaceuticals were injected, all of radiation dose rates at different distances away from patients were same with background values.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: China Medical Equipment Année: 2017 Type: Article

Documents relatifs à ce sujet