Your browser doesn't support javascript.
loading
Sterilisation effect study on granular hydroxyapatite (HA)
The Medical Journal of Malaysia ; : 85-86, 2004.
Article Dans Malayalam | WPRIM | ID: wpr-629936
ABSTRACT
Hydroxyapatite is a calcium phosphate bioceramic that has been shown by many authors to be biocompatible with bioactive properties. It is widely accepted as the best synthetic material available for surgical use as a bone graft substitute. HA granules produced by AMREC-SIRIM from local materials underwent 5 types of sterilisation techniques with different ageing periods. Samples were tested for chemical and phase composition and microbial contamination before and after being sterilised. From the microbiological tests done, none of the unsterilised positive control yielded a positive culture. Results from X-Ray diffraction studies found that all the sterilisation techniques did not chemically degrade or structurally change the HA granules significantly.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Diffraction des rayons X / Test de matériaux / Phosphates de calcium / Stérilisation / Techniques bactériologiques / Substituts osseux langue: Malayalam Texte intégral: The Medical Journal of Malaysia Année: 2004 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Diffraction des rayons X / Test de matériaux / Phosphates de calcium / Stérilisation / Techniques bactériologiques / Substituts osseux langue: Malayalam Texte intégral: The Medical Journal of Malaysia Année: 2004 Type: Article