Your browser doesn't support javascript.
loading
Intraoperative Measurement of Femorotibial Contact Pressure in Total Knee Arthroplasty / 대한정형외과학회잡지
The Journal of the Korean Orthopaedic Association ; : 149-154, 2001.
Article Dans Coréen | WPRIM | ID: wpr-649967
ABSTRACT

PURPOSE:

A new method has been devised to determine contact pressures in prosthetic components. A real time analysis of the contact pressure distribution could help fine tuned implant positioning and ligament balancing during Total Knee Arthroplasty. MATERIALS AND

METHODS:

Posterior cruciate reserving type procedures were done with externally rotated femoral cut. Femorotibial contact pressure was measured in four compartments in both flexion and extension position. A LOAD CELL located beneath the tibial plate changes the contact pressure into electric signal and DP-41S transformed the electric sign into the pressure unit. Adjunctive fine-tuning operation was done to approximate the equilibrium among the four compartments of the tibial plate.

RESULTS:

Femorotibial contact pressure were distributed between 17.1 N (Newton) to 33.4 N (average 23.6 N). A peak contact pressure was exhibited in the anteromedial compartment in extended knee and in the posterolateral compartment in flexed knee.

CONCLUSION:

A LOAD CELL electronic transducer technique was introduced. This method was tried to get the balanced equilibrium of contact pressure between Femorotibial components during TKA.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Arthroplastie / Transducteurs / Genou / Ligaments langue: Coréen Texte intégral: The Journal of the Korean Orthopaedic Association Année: 2001 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Arthroplastie / Transducteurs / Genou / Ligaments langue: Coréen Texte intégral: The Journal of the Korean Orthopaedic Association Année: 2001 Type: Article