Comparison of Conventional Chiron 348 pH/blood Gas/electrolytes Analyzer and i-STAT Portable Clinical Analyzer / 대한구급학회지
The Korean Journal of Critical Care Medicine
; : 35-40, 2000.
Article
de Ko
| WPRIM
| ID: wpr-654423
Bibliothèque responsable:
WPRO
ABSTRACT
BACKGROUND: The development of so-called "Point-of-care" (POC) devices for blood or urine analysis has resulted in many systems that are widely used at home or at the bedside. We evaluated the performance of the I-STAT portable clinical analyzer for measuring blood gases (pH, PCO2, PO2) and whole blood electrolytes (sodium, potassium and ionized calcium) with reference to a conventional blood gas analyzer (Chiron 348 pH/ blood gas/electrolytes). METHODS: Thirty samples from the arterial blood were simultaneously analyzed with I-STAT system and with Chiron 348 pH/blood gas/ electrolytes analyzer. Differences between results of two methods were analyzed by paired t-test (p<0.05). RESULTS: PO2, PCO2 and Na measured with I-STAT system and Chiron 348 PH/ blood gas/ electrolytes analyzer showed no significant differences. pH, calculated HCO3, K and Ca measured by two methods showed significant difference, but observed differences would not affect clinical decisions except ionized calcium. CONCLUSIONS: This study shows that blood gas and electrolyte analysis using the I-STAT portable device is comparable with that performed by a conventional Chiron 348 pH/blood gas/ electrolytes analyzer except ionized calcium.
Mots clés
Texte intégral:
1
Indice:
WPRIM
Sujet Principal:
Potassium
/
Calcium
/
Électrolytes
/
Gaz
/
Concentration en ions d'hydrogène
Type d'étude:
Prognostic_studies
langue:
Ko
Texte intégral:
The Korean Journal of Critical Care Medicine
Année:
2000
Type:
Article