Comparison of Impedance Cardiography with Thermodilution of Hemodynamic Parameters in Critically Ill Patients / 대한구급학회지
The Korean Journal of Critical Care Medicine
;
: 77-82, 2006.
Article
Dans Coréen
| WPRIM
| ID: wpr-656447
ABSTRACT
BACKGROUND:
To evaluate the feasibility of noninvasive hemodynamic monitoring (impedance cardiography, ICG) in critically ill patients, we compared this technique with simultaneous invasive monitoring with a pulmonary artery thermodilution catheter.METHODS:
A prospective observational study was done comparing invasive monitoring and noninvasive monitoring in 12 critically ill patients. The cardiac output (CO), the stroke volume (SV) and the systemic vascular resistance (SVR) measured by using a standard thermodilution pulmonary artery catheter technique were compared with the corresponding measurements simultaneously using an ICG.RESULTS:
The value of CO, SV and SVR measured by ICG were closely correlated to those by the thermodilution methods [r 0.659 (p<0.01), 0.536 (p<0.01), 0.738 (p<0.01)].CONCLUSIONS:
ICG can provide hemodynamic information previously available only by invasive monitoring with a thermodilution catheter.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Artère pulmonaire
/
Débit systolique
/
Thermodilution
/
Résistance vasculaire
/
Débit cardiaque
/
Cardiographie d'impédance
/
Études prospectives
/
Maladie grave
/
Impédance électrique
/
Cathéters
Type d'étude:
Étude observationnelle
Limites du sujet:
Humains
langue:
Coréen