Study on the biocompatibility of self-made nitinol alloy ventricular septal defect occluder / 第二军医大学学报
Academic Journal of Second Military Medical University ; (12)2000.
Article Dans Chinois | WPRIM | ID: wpr-677559
ABSTRACT

Objective:

To evaluate the biocompatibility of self made nitinol alloy ventricular septal defect occluder.

Methods:

Six nitinol alloy ventricular septal defect occluder were implanted in the ventricular septum by catheter in 6 normal anaesthetized open chest pigs, and the animals were observed for 45 to 120 d(2 animals).

Results:

One deaths resulted from hemorrhage and another from embolization of occluder in abdominal aorta during the placement procedure. Successful placement of the occluder was achieved in 4 animals. Four animals were killed at 45, 60 and 120 d. Postmortem gross and microscopic examination of 4 devices 45 to 120 d after placement showed that both the right and left ventricular discs of the occluder were completely covered by a smooth, shiny, glistening thin layer of neoendocardium, and the surface of neoendocardium was covered by a monolayer of endothelium like cells. The inflammatory infiltrate around the occluder was found at 45 d, and disappeared and fibrosis formed at 120 d. These appeared as a repair process of the injury. Embolization in lung,liver, spleen, kidney, intestinal and colon were not found.

Conclusion:

These suggest that the self made nitinol alloy ventricular septal defect occluder has good biocompatibility. [

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Academic Journal of Second Military Medical University Année: 2000 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Academic Journal of Second Military Medical University Année: 2000 Type: Article