Application risk and management strategy of infusion pumps / 医疗卫生装备
Chinese Medical Equipment Journal
; (6): 138-141, 2017.
Article
de Zh
| WPRIM
| ID: wpr-699925
Bibliothèque responsable:
WPRO
ABSTRACT
Objective To analyze the risk factors of the infusion pump to provide evidence-based basis for its safety management and monitoring.Methods Totally 1 000 pieces of questionnaires were issued for 100 infusion pumps from 9 grade-A tertiary hospitals in Beijing.The abnormalities and the correlation between the indexes were analyzed for the infusion pump,the indexes were compared on their effects on the abnormalities.Results Statistical analysis on data samples showed that the main factors included the manufacturer of infusion pump and pipeline,service time of infusion pump,compatibility between infusion pump and pipeline as well as the user department.Conclusion The management facility has to pose emphases on safety and utilization management as well as standardized application of the infusion pipeline so as to reduce adverse events efficiently.
Texte intégral:
1
Indice:
WPRIM
Type d'étude:
Etiology_studies
/
Risk_factors_studies
langue:
Zh
Texte intégral:
Chinese Medical Equipment Journal
Année:
2017
Type:
Article