Experimental study of dosimetry for CTLD-J4000 energy differential type personal dosimeter / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 374-378, 2018.
Article Dans Chinois | WPRIM | ID: wpr-708072
ABSTRACT
Objective To understand the dosimetric performance of energy differential type dosimeter.Methods A differential type dosimeter (CTLD-J4000) was chosed to conduct a series of dosimetric performance tests,including linear standard curves,relative response to mean photon radiation energy and incidence angle.Results The linear correlation of CTLD-J4000 energy differential dosimeter was good,with the average photon radiation energy response deviation within ± 20% and the maximum incidence angle response deviation of 10%.The energy identification could be achieved according to the energy reference value provided by the manufacturer to a certain extent.Conclusions Further experiments are needed to find out the factors influencing measurement result from different dose values and incidence angles.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Type d'étude: Étude pronostique langue: Chinois Texte intégral: Chinese Journal of Radiological Medicine and Protection Année: 2018 Type: Article