AG200 and polysomnography in patients with obstructive sleep apnea-hypopnea syndrome / 临床耳鼻咽喉头颈外科杂志
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
;
(24): 1691-1693, 2014.
Article
Dans Chinois
| WPRIM
| ID: wpr-747661
ABSTRACT
OBJECTIVE@#To study the clinical value of polysomnography (PSG) and ApneaGraph (AG200) in the diagnosis evaluation of obstructive sleep apnea hypopnea syndrome (OSAHS).@*METHOD@#From January to December in 2012, 67 OSAHS patients diagnosed by PSG were examined by AG200. The apnea hypopnea index (AHI), hypopnea index (HI), apnea index (AI), the lowest oxygen saturation (LSaO2) was were detected and the results were analyzed statistically.@*RESULT@#Significant differences were observed in AI, AHI, LSaO2 between AG200 and PSG (P 0.05). AHI, HI and LSaO2 was were significantly correlated between AG200 and PSG (r = 0.870, 0.743, 0.374, 0.716, P < 0.01).@*CONCLUSION@#AG200 could not replace PSG but could identify the level of upper airway obstruction.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Polysomnographie
/
Syndrome d'apnées obstructives du sommeil
/
Diagnostic
/
Méthodes
Type d'étude:
Etude diagnostique
Limites du sujet:
Femelle
/
Humains
/
Mâle
langue:
Chinois
Texte intégral:
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
Année:
2014
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS