Your browser doesn't support javascript.
loading
Accuracy and Efficacy of Impedance Cardiography as a Non-Invasive Cardiac Function Monitor
Yonsei Medical Journal ; : 735-741, 2019.
Article Dans Anglais | WPRIM | ID: wpr-762112
ABSTRACT

PURPOSE:

The most common method of monitoring cardiac output (CO) is thermodilution using pulmonary artery catheter (PAC), but this method is associated with complications. Impedance cardiography (ICG) is a non-invasive CO monitoring technique. This study compared the accuracy and efficacy of ICG as a non-invasive cardiac function monitoring technique to those of thermodilution and arterial pressure contour. MATERIALS AND

METHODS:

Sixteen patients undergoing liver transplantation were included. Cardiac index (CI) was measured by thermodilution using PAC, arterial waveform analysis, and ICG simultaneously in each patient. Statistical analysis was performed using intraclass correlation coefficient (ICC) and Bland-Altman analysis to assess the degree of agreement.

RESULTS:

The difference by thermodilution and ICG was 1.13 L/min/m², and the limits of agreement were −0.93 and 3.20 L/min/m². The difference by thermodilution and arterial pressure contour was 0.62 L/min/m², and the limits of agreement were −1.43 and 2.67 L/min/m². The difference by arterial pressure contour and ICG was 0.50 L/min/m², and the limits of agreement were −1.32 and 2.32 L/min/m². All three percentage errors exceeded the 30% limit of acceptance. Substantial agreement was observed between CI of thermodilution with PAC and ICG at preanhepatic and anhepatic phases, as well as between CI of thermodilution and arterial waveform analysis at preanhepatic phase. Others showed moderate agreement.

CONCLUSION:

Although neither method was clinically equivalent to thermodilution, ICG showed more substantial correlation with thermodilution method than with arterial waveform analysis. As a non-invasive cardiac function monitor, ICG would likely require further studies in other settings.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Artère pulmonaire / Thermodilution / Débit cardiaque / Cardiographie d'impédance / Transplantation hépatique / Impédance électrique / Cathéters / Pression artérielle / Méthodes Limites du sujet: Humains langue: Anglais Texte intégral: Yonsei Medical Journal Année: 2019 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Artère pulmonaire / Thermodilution / Débit cardiaque / Cardiographie d'impédance / Transplantation hépatique / Impédance électrique / Cathéters / Pression artérielle / Méthodes Limites du sujet: Humains langue: Anglais Texte intégral: Yonsei Medical Journal Année: 2019 Type: Article