Design and Implementation of Automated QA Software for Cone Beam CT Image / 中国医疗器械杂志
Zhongguo Yi Liao Qi Xie Za Zhi
; (6): 25-28, 2019.
Article
de Zh
| WPRIM
| ID: wpr-772574
Bibliothèque responsable:
WPRO
ABSTRACT
In order to improve the speed and accuracy of cone beam CT image quality assurance, an automatic image quality assurance software based on feature extraction is designed and implemented. This paper introduces how the software can realize the positioning of Catphan500 phantoms and the selection of ROI in each module through Canny algorithm and Hough transform circle detection. For the different modules in the Catphan500 phantom, this paper describes the calculation methods of HU accuracy, spatial linearity, HU uniformity and spatial linearity in the software. Finally, the article verifies the feasibility of the software through experiments.
Mots clés
Texte intégral:
1
Indice:
WPRIM
Sujet Principal:
Contrôle de qualité
/
Automatisation
/
Algorithmes
/
Logiciel
/
Fantômes en imagerie
/
Tomodensitométrie à faisceau conique