Your browser doesn't support javascript.
loading
Acceptance Test and Image Quality Assurance of MRI Simulator Equipment / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 455-459, 2018.
Article Dans Chinois | WPRIM | ID: wpr-775529
ABSTRACT
MRI simulator(MRI-Sim) images have unique clinical advantages with higher resolution of soft tissue and clearer visualization of tissue boundaries. Thus, the precise positioning of the tumor target area can be achieved and it is widely used in the field of radiotherapy. This article focuses on the acceptance test project and image quality assurance work of MRI-Sim equipment. The obtained ACR phantom images were used to analyze various image quality assurance indicators, and the results all reached the set standards, thereby ensuring that the obtained images meet the requirements of clinical applications.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Assurance de la qualité des soins de santé / Imagerie par résonance magnétique / Fantômes en imagerie Type d'étude: Guide de pratique langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2018 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Assurance de la qualité des soins de santé / Imagerie par résonance magnétique / Fantômes en imagerie Type d'étude: Guide de pratique langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2018 Type: Article