Performance Evaluation of MEDITAPE UC-11A Strip Test in Estimating the Urine Albumin-to-Creatinine Ratio and Urine Protein-to-Creatinine Ratio
Laboratory Medicine Online
;
: 52-57, 2020.
Article
Dans Anglais
| WPRIM
| ID: wpr-782247
ABSTRACT
30 mg/g), the concordance rate, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of UACR, analyzed using MEDITAPE UC-11A, were 80.5, 97.5, 67.0, 70.3, and 97.1%, respectively. Using UPCR, analyzed via quantitative assay, as a reference to estimate proteinuria (UPCR >0.15 g/g), the concordance rate, sensitivity, specificity, PPV, and NPV of UPCR, analyzed using MEDITAPE UC-11A, were 86.7, 94.4, 81.5, 77.6, and 95.6%, respectively.CONCLUSIONS:
UACR and UPCR, analyzed using MEDITAPE UC-11A, exhibited relatively high sensitivity and NPV, which is beneficial for laboratory screening for both albuminuria and proteinuria.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Protéinurie
/
Dépistage de masse
/
Maladie chronique
/
Sensibilité et spécificité
/
Albuminurie
/
Insuffisance rénale chronique
/
Hypertension artérielle
/
Maladies du rein
Type d'étude:
Etude diagnostique
/
Étude de dépistage
Limites du sujet:
Humains
langue:
Anglais
Texte intégral:
Laboratory Medicine Online
Année:
2020
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS