Your browser doesn't support javascript.
loading
Study on the effect of a cold environment on the quality of three video laryngoscopes: McGrath MAC, GlideScope Ranger, and Pentax Airway Scope
Clinical and Experimental Emergency Medicine ; (4): 351-355, 2019.
Article Dans Anglais | WPRIM | ID: wpr-785626
ABSTRACT

OBJECTIVE:

Several environmental factors influence the prehospital use of video laryngoscopes (VLs). For example, fogging of the VL lens can occur in cold environments, and the low temperature can cause the VLs to malfunction. As relevant research on the effect of environment on VLs is lacking, we aimed to study the effect of a cold environment on three commonly used VLs.

METHODS:

McGrath MAC, Pentax Airway Scope (AWS), and GlideScope Ranger were exposed to temperatures of -5°C, -10°C, -20°C, and -25°C for 1 hour each and then applied to a manikin in a thermohydrostat room 5 times. Immediately after turning on the power and inserting the blade, the time until an appropriate glottic image appeared on the screen was measured.

RESULTS:

McGrath MAC was able to accomplish immediate intubation regardless of the temperature drop. However, GlideScope Ranger required an average of 4.9 seconds (-5°C to -20°C) and 10.1 seconds (-25°C) until appropriate images were obtained for intubation. AWS showed adequate image acquisition immediately after blade insertion despite slight fogging at -20°C, but at -25°C, images suitable for intubation did not appear on the screen for an average of 4.7 minutes.

CONCLUSION:

All three devices appear to be usable without any limitations up to -20°C. However, GlideScope Ranger and AWS may not produce images immediately at temperatures below -25°C. Thus, medical practitioners performing VL in a cold environment should be aware of the characteristics of the VL devices in advance.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Temps (météorologie) / Basse température / Laryngoscopes / Intubation / Intubation trachéale / Mannequins langue: Anglais Texte intégral: Clinical and Experimental Emergency Medicine Année: 2019 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS