Verification of doses to PTV and OAR and 2D dose distribution in IMRT for 7 accelerators in Sichuan province / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 145-149, 2020.
Article Dans Chinois | WPRIM | ID: wpr-799421
ABSTRACT
Objective@#To study a method for verifying the doses to PTV and OAR as well as the 2D dose distribution arising from IMRT through using radiochromic films and TLDs.@*Methods@#Totally 7 medical electronic linear accelerators from Varian, Siemens and Elekta were selected. The polystyrene phantom provided by IAEA was conducted with CT scan. After irradiation with 6 MV X-rays, the TLDs and films were returned to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.@*Results@#According to the IAEA requirements, the relative deviations between TLD-measured and TPS-planned values for PTV and OAR doses were both within ±7.0%. For PTV, the measured relative deviations for 5 accelerators were in the range of -4.0% to 3.4%, consistent with the IAEA requirements, whereas the values for the other 2 accelerators were in the range of -7.0% to 10.6%, not consistent with the requirements. For OAR, the values for 4 accelerators were in the range of -5.6% to 3.3%, consistent with the IAEA requirements, whereas the values for the other 3 accelerators were in the range of -20.8% to 11.5%, not meeting the requirements. As required by the IAEA, the 2D dose distribution 3 mm/3% pass rate should be higher than 90%. The measured values for 5 accelerators were in the range of 91.8% to 98.5%, consistent with the requirements, whereas the values measured for the other 2 were 45.0% and 77.0% respectively, not meeting the requirements.@*Conclusions@#It is feasible for using TLDs and radiochromic films to verify the doses to PTV and OAR and the 2D dose distribution in IMRT. This method should be applied to not only quality verification but also hospital internal audit to the extent possible.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Radiological Medicine and Protection Année: 2020 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Radiological Medicine and Protection Année: 2020 Type: Article