On the consistency of different methods for detecting aldosterone concentration in blood / 中华内分泌代谢杂志
Chinese Journal of Endocrinology and Metabolism ; (12): 934-938, 2019.
Article Dans Chinois | WPRIM | ID: wpr-800752
ABSTRACT
Objective@#To evaluate the consistency of different methods for detecting aldosterone concentration in blood and to establish a reference interval of serum aldosterone concentration by liquid chromatography-tandem mass spectrometry(LC-MS/MS).@*Methods@#Concentrations of blood aldosterone were measured by LC-MS/MS, chemiluminescent assays (Diasorin, Domestic A and B systems) and radioimmunoassay (RIA) in 138 healthy adults, 67 patients with essential hypertension and 23 patients with primary aldosteronism.@*Results@#Aldosterone concentrations measured by various methods were quiet different(P<0.01). Spearman correlation analysis showed that the correlation coefficient were 0.776(P<0.01) between CLIA (Diasorin) and LC-MS/MS, 0.805(P<0.01) between CLIA (Domestic A) and LC-MS/MS, 0.058(P>0.05) between CLIA(Domestic B) and LC-MS/MS, 0.338(P<0.01) as well as between RIA and LC-MS/MS. Bland-Altman analysis for the measurements showed poor consistency among methods for aldosterone concentrations. The reference interval was 15.2-222.2 pg/ml for serum aldosterone concentrations by LC-MS/MS.@*Conclusions@#There are significant differences of blood aldosterone concentrations determined by different methods. Clinically, a specified reference interval might be needed while using different methods.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Endocrinology and Metabolism Année: 2019 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Endocrinology and Metabolism Année: 2019 Type: Article