Your browser doesn't support javascript.
loading
Development of a force balance telemetry device for total knee arthroplasty / 医用生物力学
Journal of Medical Biomechanics ; (6): E112-E117, 2010.
Article Dans Chinois | WPRIM | ID: wpr-803655
ABSTRACT
Objective A new type of tibial femoral force balancing telemetry device was developed for measuring force balance on the tibial femoral contact surface in knee joint. MethodThe force balance telemetry device was designed imitating the configuration of tibial part of total knee prosthesis, in which force sensors and telemetry circuit were embedded. During total knee arthroplasty(TKA), this device was implanted as a tibial prosthesis to measure tibial femoral force while the knee joint was being placed at different angles. The tension balance was adjusted by releasing soft tissue around knee joint. ResultsThe results indicat that this device has advantage of high accuracy for force measurement (r>0.98,RMS= 65 g), which is sufficient to meet clinical demand. ConclusionsThe measurement of force balance on the tibial femoral contact surface with telemetry device can provide new insight into the estimation of optimal knee prosthesis position and selection of appropriate polyethylene insert.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Journal of Medical Biomechanics Année: 2010 Type: Article

Documents relatifs à ce sujet