Dosimetric verification of flattening filter free model based on TrueBeam accelerator using ArcCheck system / 中南大学学报(医学版)
Journal of Central South University(Medical Sciences)
;
(12): 864-868, 2018.
Article
Dans Chinois
| WPRIM
| ID: wpr-813183
ABSTRACT
To study the feasibility of ArcCheck verification system in dosimetric verification for stereotactic radiotherapy (SRT) the stereotactic radiotherapy (SRT) with flattening filter free (FFF) model.
Methods:
A total of 76 cases under SRT treatment plans were introduced into ArcCheck phantom and recalculated. Threshold criteria was set as (3%, 3 mm, 10%) or (2%, 2 mm, 10%). The calculated dose distribution and the measured dose distribution of ArcCheck phantom were compared by means of distance to agree (DTA) and Gamma analysis method respectively.Results:
Based on the threshold criteria (3%, 3 mm, 10%), the relative and absolute mean pass rates of SRT treatment plans by DTA and Gamma analysis were greater than 95%. Based on the threshold criteria (2%, 2 mm, 10%), the relative and absolute mean pass rates of SRT treatment plan by DTA and Gamma analysis were about 90%. The dose pass rate of Gamma analysis method was slightly higher than that of DTA analysis method (P<0.001).Conclusion:
The ArcCheck verification system is a rapid and accurate method for SRT dose verification, and discrepancies are found in different analysis methods.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Dosimétrie en radiothérapie
/
Planification de radiothérapie assistée par ordinateur
/
Études de faisabilité
/
Radiochirurgie
/
Fantômes en imagerie
/
Radiothérapie conformationnelle avec modulation d'intensité
/
Méthodes
Type d'étude:
Étude pronostique
Limites du sujet:
Humains
langue:
Chinois
Texte intégral:
Journal of Central South University(Medical Sciences)
Année:
2018
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS