Postural Intraocular Pressure Change at Trendelenberg Position Measured by Rebound Tonometer
Journal of the Korean Ophthalmological Society ; : 247-251, 2014.
Article Dans Coréen | WPRIM | ID: wpr-90229
ABSTRACT

PURPOSE:

This study was carried out to evaluate the postural intraocular pressure (IOP) change in Trendelenburg, reverse Trendelenburg, and supine positions in healthy young males.

METHODS:

We measured the IOP values of 5 healthy young male volunteers (10 eyes) using an Icare PRO rebound tonometer in sitting, Trendelenburg, reverse Trendelenburg, and supine positions.

RESULTS:

The mean IOP in the supine position (18.63 mm Hg) was significantly higher (p < 0.01) than in the sitting position (15.31 mm Hg). When maintaining the Trendelenburg position, IOP gradually increased.

CONCLUSIONS:

In our study, the Trendelenburg position significantly increased the IOP compared to that in the supine position. The effects of increased IOP should be considered in situations that require Trendelenburg positioning, such as exercise or surgery.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Bénévoles / Décubitus dorsal / Position déclive / Pression intraoculaire / Dextriferron Limites du sujet: Humains / Mâle langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2014 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Bénévoles / Décubitus dorsal / Position déclive / Pression intraoculaire / Dextriferron Limites du sujet: Humains / Mâle langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2014 Type: Article