Removal torque of a two-piece zirconia abutment with a novel titanium component in an internal connection implant system after dynamic cyclic loading / 대한치과보철학회지
The Journal of Korean Academy of Prosthodontics
; : 151-155, 2017.
Article
de En
| WPRIM
| ID: wpr-91672
Bibliothèque responsable:
WPRO
ABSTRACT
PURPOSE: The aim of this study was to evaluate screw removal torque of the two-piece zirconia abutment with the novel titanium component compared to the conventional onepiece titanium abutment in the internal connection implant before and after dynamic cyclic loading. MATERIALS AND METHODS: Two types of the abutment assemblies with internal connection were prepared and divided into the groups; titanium abutment-titanium abutment screw assemblies as control, and zirconia abutment-titanium socket-titanium abutment screw assemblies as experimental group. A total of 12 abutments and implants were used of six assemblies each group. Each assembly was tightened to 30 Ncm. A cyclic load of 300 N at an angle of 30 degrees in reference to the loading axis was applied until one million cycles or failure. The removal torque values (RTVs) of the abutment screws were measured with a digital torque gauge before and after cyclic loading. The RTVs of the pre load and post load were analyzed with t-test, and P-values .05). CONCLUSION: Within the limitation of this study, the two-piece zirconia abutment with the titanium component did not show a significant RTV difference of the abutment screw compared to the titanium abutment after dynamic cyclic loading.
Mots clés
Texte intégral:
1
Indice:
WPRIM
Sujet Principal:
Titane
/
Moment de torsion
langue:
En
Texte intégral:
The Journal of Korean Academy of Prosthodontics
Année:
2017
Type:
Article