A Prospective Observational Clinical and Radiological Study of a Modular Bridging External Fixator for Unstable Distal Radius Fractures
Malaysian Orthopaedic Journal
;
: 108-114, 2021.
Article
Dans Anglais
| WPRIM
| ID: wpr-923066
ABSTRACT
@#Introduction:
Distal Radius Fractures (DRFs), with a reported annual incidence of 600,000, are common injuries treated by trauma surgeons. This prospective observational study aims to assess the efficacy of a modular external fixation system in the treatment of unstable distal radius fractures at 12-months follow-up. Materials andmethods:
Between December 2014 and December 2016, 35 patients (female 21, male14; mean age 62.5), with unstable DRFs, treated with modular external fixation system, were selected for this prospective observational study. All the patients underwent clinical and radiological reviews at follow-up.Results:
At 12-month follow-up, a mean DASH score of 15.73 and a mean PRWE score 20.10 were recorded. Mean radial inclination was 19.92°; mean ulnar variance was 1.12 mm and mean palmar inclination was 9.76°.Conclusion:
Modular external fixator system revealed clinically and radiologically effective in the treatment of unstable and comminuted DRFs. Additional K-wires should be used to complement the fracture fixation, when there is unacceptable fragment reduction only with external fixator.