Your browser doesn't support javascript.
loading
Simulation Study on the Influence of Sampling Delay on the Accuracy of Energy Metabolism Measurement / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 382-387, 2022.
Article Dans Chinois | WPRIM | ID: wpr-939752
ABSTRACT
Indirect energy metabolism measurement is the gold standard for providing nutritional support for critical illness. The accuracy of the measurement data directly affects the outcome of the disease. In order to study the influence of sampling delay on the accuracy of energy metabolism measurement under mechanical ventilation, the Matlab/Simulink platform and respiratory electrical model were used for simulation and quantitative analysis. The results show that the error of indirect energy metabolism measurement increases with the increase of sampling delay, the error of sampling delay in mechanical ventilation mode is larger than that of spontaneous breathing, and the error of sampling delay in PCV mode of mechanical ventilation is larger than that in VCV mode. Therefore, there should be different sampling delay compensation strategies under severe mechanical ventilation and its different control modes.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Ventilation artificielle / Simulation numérique / Maladie grave / Métabolisme énergétique Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2022 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Ventilation artificielle / Simulation numérique / Maladie grave / Métabolisme énergétique Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2022 Type: Article