Experimental Study of MAR Algorithm in Metal Artifact Removal of CT Simulator / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 24-27, 2020.
Article Dans Chinois | WPRIM | ID: wpr-942690
ABSTRACT
OBJECTIVE@#To explore the application value of MAR algorithm in metal artifact removal of CT simulator.@*METHODS@#CT phantom with titanium plate was scanned using conventional algorithms and MAR algorithms, respectively. Artifact index(AI), contrast-to-noise ratio(CNR) and AI values at different slices were used to analyze the artifact images.@*RESULTS@#In artifact index, MAR algorithm (10.28±2.60) is significantly lower than conventional algorithm (20.65±5.04); In contrast-to-noise ratio index, MAR algorithm (7.81±1.12) is better than conventional algorithm (5.61±1.36). The above indicators were statistically significant in both algorithms (P<0.01). In the slices affected by metal artifacts, the artifact index decreased by 21.72%~88.40% after the MAR algorithm.@*CONCLUSIONS@#MAR algorithm can significantly reduce the metal artifacts and improve the clinical value of CT data.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Titane / Algorithmes / Tomodensitométrie / Artéfacts / Fantômes en imagerie / Métaux Type d'étude: Étude pronostique langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2020 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Titane / Algorithmes / Tomodensitométrie / Artéfacts / Fantômes en imagerie / Métaux Type d'étude: Étude pronostique langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2020 Type: Article