Research on 7.0 T Magnetic Resonance Based Electrical Properties Tomography Based on Bloch-Siegert Shift / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 189-193, 2020.
Article
Dans Chinois
| WPRIM
| ID: wpr-942725
ABSTRACT
Magnetic resonance based electrical properties tomography (MREPT) is a different method from proton density imaging, Bloch-Siegert shift (BSS) is used in this paper to reconstruct the radiofrequency (RF) field amplitude and calculate the distribution of the permittivity constant. The phase of the RF field is approximated by the phase component of the magnetization intensity, and the conductivity distribution is calculated. In the experiment, Bruker 7.0 T magnetic resonance device was used to image two water models and in vivo Balb/c mice to obtain the image of electrical characteristics. Experimental results show that the Bloch-siegert B1+ image is significantly more efficient than the double-angle B1+ image. The study can provide a reference for selecting appropriate B1 mapping technology for B1 field imaging of electrical characteristics organizations, and provide basic research support for promoting the practical application of magnetic resonance characteristics.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Algorithmes
/
Imagerie par résonance magnétique
/
Spectroscopie par résonance magnétique
/
Tomographie
/
Fantômes en imagerie
Type d'étude:
Étude pronostique
Limites du sujet:
Animaux
langue:
Chinois
Texte intégral:
Chinese Journal of Medical Instrumentation
Année:
2020
Type:
Article