ABSTRACT
This study mainly discusses the contents, methods and characteristics of the collection of operation and maintenance data, as well as the establishment and evaluation methods of the distribution model of the failure time of medical electrical equipment. The distribution models of failure time at three levels of medical electrical equipment are established by linear regression method and goodness of fit testThe first is the device level MTBF distribution model, the second is the failure rate distribution model of the failure mode of key components, the third is the calculation model of the influence coefficient of influence factor on the failure mode of key components. This study presents a method of establishing MTBF segment model and implements a calculation model of influence coefficient varying with time.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Maintenance Type d'étude: Étude pronostique langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2020 Type: Article