Normalization and parallel comparison of the detection results of antibodies against SARS-CoV-2 / 中华微生物学和免疫学杂志
Chinese Journal of Microbiology and Immunology
;
(12): 761-768, 2022.
Article
Dans Chinois
| WPRIM
| ID: wpr-958254
ABSTRACT
Objective:
To improve the consistency of test results through reducing inter-laboratory variation in SARS-CoV-2 antibody detection with WHO SARS-CoV-2 antibody candidate international standard (IS, sample G) and antibody reference panel (samples E, F, H, I, J).Methods:
Ten WHO samples (A-J) including the candidate IS and reference panel were evaluated using different methods, such as microneutralization tests based on live SARS-CoV-2, pseudovirus neutralization assay and commercial ELISA kits. The test results were compared using statistical analysis.Results:
Using IS (sample G) as a reference, the relative concentrations of other samples could be determined with less variation. ELISA and pseudovirus neutralization assay had consistent results with those obtained with the microneutralization test based on SARS-COV-2 strain HB02. Weakly positive samples could be detected only by a certain kit.Conclusions:
The availability of an IS for antibodies would facilitate the standardization of SARS-CoV-2 antibody detection methods. The reference panel fitted all the assays based on the SARS-CoV-2 prototype Wuhan strain. Pseudovirus neutralization assay and ELISA could be used as alternatives to live SARS-CoV-2-based neutralization test to some extent.