Resultados visuales en la corrección de la alta miopía con implante de lente fáquica ACR-128 / Visual results of correction of high myopia with ACR-128 phakic lens implantation
Cárdenas Díaz, Taimi; Qi Li, Feng; Pérez Suárez, Raúl; Cruz Izquierdo, Dunia; Monteagudo Hernández, Ketty; Cabrera Ruiz, Yaima.
  • Cárdenas Díaz, Taimi; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Qi Li, Feng; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Pérez Suárez, Raúl; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Cruz Izquierdo, Dunia; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Monteagudo Hernández, Ketty; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Cabrera Ruiz, Yaima; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
Rev. cuba. oftalmol ; 32(2): e611, abr.-jun. 2019. tab
Artigo em Espanhol | LILACS | ID: biblio-1093686
RESUMEN
RESUMEN

Objetivo:

Describir los resultados visuales en la corrección de la alta miopía con implante de lente fáquica ACR-128.

Métodos:

Se realizó un estudio descriptivo, observacional longitudinal y prospectivo en 60 ojos de 32 pacientes con miopía corregida con lente fáquica ACR-128. Se determinó el componente esférico esperado y observado, el cilindro queratométrico, las agudezas visuales sin corrección y mejor corregidas en el pre y en el posoperatorio y el astigmatismo inducido. El análisis estadístico se realizó con la Prueba T para datos pareados, con una significación del 95 por ciento.

Resultados:

La edad media fue 27,41 ± 5,91 años, el equivalente esférico preoperatorio -11,54 ± 3,20 dioptrías y el 68,75 por ciento eran femeninas. El componente esférico en dioptrías esperado (-0,53 ± 0,37) y observado (-0,42 ± 0,47) sin diferencias (p= 0,0742). Entre ± 1,00 el 91,67 por ciento y entre ± 0,50 el 70 por ciento. Ningún ojo quedó por encima de +0,50 dioptrías. El cilindro queratométrico en dioptrías, pre (1,44 ± 0,76) y posoperatorio (1,49 ± 0,84) sin astigmatismo inducido (p= 0,6377). El 100 por ciento tenía agudeza visual sin corrección preoperatoria ≤ 0,1 y posoperatoria ≥ 0,3. Después de la cirugía el 10 por ciento alcanzaba 1,0 y 71,6 por ciento ≥ 0,5. Solo el 28,33 por ciento tenía la unidad en el preoperatorio, y el 70 por ciento en el posoperatorio (98,33 por ciento ≥ 0,7).

Conclusiones:

El implante de lente fáquica ACR-128 proporciona corrección refractiva y predictibilidad favorables para el paciente, al reducir el componente esférico al deseado, no inducir astigmatismo y mejorar la agudeza visual, todo lo que se traduce en un adecuado resultado visual(AU)
ABSTRACT
ABSTRACT

Objective:

Describe the visual results of correction of high myopia with ACR-128 phakic lens implantation.

Methods:

An observational descriptive longitudinal prospective study was conducted of 60 eyes of 32 patients with myopia corrected with the ACR-128 phakic lens. Determination was made of the expected and observed spherical component, the keratometric cylinder, uncorrected and best corrected visual acuity in the pre- and postoperative periods, and induced astigmatism. Statistical analysis was based on the paired T-test with a significance level of 95 percent.

Results:

Mean age was 27.41 ± 5.91 years, preoperative spherical equivalent was -11.54 ± 3.20 diopters, and 68.75 percent of the patients were female. Spherical component in diopters expected (-0.53 ± 0.37) and observed (-0.42 ± 0.47) without differences (p= 0.0742). Between ± 1.00 diopters 91.67 percent and between ± 0.50 diopter 70 percent. No eye was above +0.50D. Keratometric cylinder in diopters, preoperative (1.44 ± 0.76) and postoperative (1.49 ± 0.84) without induced astigmatism (p=0.6377). In 100 percent visual acuity without correction ≤ 0.1 preoperative and ≥ 0.3 postoperative. After surgery 10 percent reached 1.0 and 71.6 percent ≥ 0.5. Only 28.33 percent had the unit corrected in the preoperative period and 70 percent in the postoperative period (98.33 percent ≥ 0.7).

Conclusions:

ACR-128 phakic lens implantation provides patients with favorable refractive correction and predictability. This is achieved by reducing the spherical component to desired values, not inducing astigmatism and improving visual acuity, all of which leads to an adequate visual result(AU)
Assuntos

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Procedimentos Cirúrgicos Refrativos / Lentes Intraoculares Fácicas / Miopia Tipo de estudo: Estudo observacional / Fatores de risco Limite: Adulto / Feminino / Humanos Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2019 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología "Ramón Pando Ferrer"/CU

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Procedimentos Cirúrgicos Refrativos / Lentes Intraoculares Fácicas / Miopia Tipo de estudo: Estudo observacional / Fatores de risco Limite: Adulto / Feminino / Humanos Idioma: Espanhol Revista: Rev. cuba. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2019 Tipo de documento: Artigo País de afiliação: Cuba Instituição/País de afiliação: Instituto Cubano de Oftalmología "Ramón Pando Ferrer"/CU