Your browser doesn't support javascript.
loading
Confiabilidade intra e interexaminador da algometria para mensurar o limiar da dor em idosos institucionalizados / Intra and inter-rater reliability of algometry to measure pain threshold in institutionalized elderly
Prestes, Yandra Alves; Souza, Marilene dos Santos; Pereira, Mayara Souza Pereira; Ervati, Rosiane; Campos, Hércules Lázaro Morais.
  • Prestes, Yandra Alves; Universidade Federal do Amazonas. Coari. BR
  • Souza, Marilene dos Santos; Universidade Federal do Amazonas. Coari. BR
  • Pereira, Mayara Souza Pereira; 0000-0002-1498-8668. Coari. BR
  • Ervati, Rosiane; Universidade Federal do Espírito Santo. Vitória. BR
  • Campos, Hércules Lázaro Morais; Universidade Federal do Amazonas. Coari. BR
Rev. Pesqui. Fisioter ; 10(3): 451-460, ago.2020. ilus, tab
Artigo em Inglês, Português | LILACS | ID: biblio-1223933
RESUMO
A dor passa ser um fator importante quando se estuda a população idosa, pois apresenta um grande impacto nas atividades diárias, aumentando as comorbidades e a mortalidade, transformando-se em um sério problema de saúde pública.

OBJETIVOS:

Avaliar a confiabilidade das medidas intraexaminador e interexaminador de um algômetro digital em mensurar o limiar da dor no quadril de idosos institucionalizados e frágeis.

MATERIAIS E MÉTODOS:

A amostra foi de forma intencional de acordo com a capacidade cognitiva dos idosos para responderem às avaliações; dois examinadores realizaram a algometria na região do quadril em 17 idosos residentes de uma instituição de longa permanência.

RESULTADOS:

Os coeficientes de correlação intraclasse (CCI) demonstraram alta confiabilidade do instrumento intra-examinador - iguais a 0,877 e muito alta interexaminador - iguais a 0,973.

CONCLUSÃO:

Há confiabilidade no algômetro digital em mensurar o limiar de dor no quadril em idosos de uma instituição de longa permanência.
ABSTRACT
Pain becomes an important factor when studying the elderly population, as it has a great impact on daily activities, increasing comorbidities and mortality, becoming a serious public health problem.

OBJECTIVES:

To evaluate the reliability of the intra-examiner and inter-examiner measurements of a digital algometer in measuring the hip pain threshold of institutionalized and fragile elderly people. MATERIALS AND

METHODS:

The sample was intentionally in accordance with the elderly's cognitive ability to respond to assessments; two examiners performed algometry in the hip region on 17 elderly residents of a long-term institution.

RESULTS:

The intraclass correlation coefficients (ICC) high reliability of the intraexaminer instrument - equal to 0.877 and very high interexaminer - equal to 0.973.

CONCLUSION:

There is reliability in the digital algometer in measuring the hip pain threshold in the elderly in a long-term institution.
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Dor Idioma: Inglês / Português Revista: Rev. Pesqui. Fisioter Assunto da revista: Farmacologia / Terapˆutica / Toxicologia Ano de publicação: 2020 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: 0000-0002-1498-8668/BR / Universidade Federal do Amazonas/BR / Universidade Federal do Espírito Santo/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Dor Idioma: Inglês / Português Revista: Rev. Pesqui. Fisioter Assunto da revista: Farmacologia / Terapˆutica / Toxicologia Ano de publicação: 2020 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: 0000-0002-1498-8668/BR / Universidade Federal do Amazonas/BR / Universidade Federal do Espírito Santo/BR