Utilización de un dispositivo móvil de tecnología electrónica para tamizaje de fibrilación auricular: estudio piloto / Use of a mobile electronic technology device for atrial fibrillation screening: pilot study / Uso de dispositivo de tecnologia eletrônica móvel para rastreamento de fibrilação atrial: estudo piloto
Estragó Merola, Virginia Sandra; Muñoz Wolf, Matías; Álvarez Vaz, Ramón Fernando; Reyes Cabrera, María Ximena; Reyes Caorsi, Walter Jorge.
Afiliação
  • Estragó Merola, Virginia Sandra; Comisión Honoraria para la Salud Cardiovascular. Montevideo. UY
  • Muñoz Wolf, Matías; Comisión Honoraria para la Salud Cardiovascular. Montevideo. UY
  • Álvarez Vaz, Ramón Fernando; Universidad de la República. Facultad de Ciencias Económicas y de Administración. Montevideo. UY
  • Reyes Cabrera, María Ximena; Comisión Honoraria para la Salud Cardiovascular. Montevideo. UY
  • Reyes Caorsi, Walter Jorge; Comisión Honoraria para la Salud Cardiovascular. Montevideo. UY
Rev. urug. cardiol ; 36(2): e201, ago. 2021. ilus, tab
Article em Es | LILACS, UY-BNMED, BNUY | ID: biblio-1289996
Biblioteca responsável: UY1.1
RESUMEN
La fibrilación auricular (FA) es un problema de salud pública que genera mortalidad y morbilidad, con su mayor impacto en mayores de 65 años. Su detección es especialmente relevante para la población de riesgo. Este estudio piloto propone valorar la utilidad de un dispositivo móvil de tecnología electrónica (DMTE) para el tamizaje de FA. Objetivo primario evaluar la validez y confiabilidad de un DMTE para identificar FA. Objetivo secundario validar los procesos de recolección, transmisión, almacenamiento, procesamiento e interpretación de los datos obtenidos.

Método:

los asistentes a 5 talleres del Plan Ibirapitá (entre el 15 de octubre y el 30 de noviembre de 2018) fueron invitados a participar del estudio. Se utilizó un DMTE (sensor KardiaMobile de AliveCor®) cuyo registro se contrastó con el trazado del ECG de 12 derivaciones realizado en forma casi simultánea. Se compararon los diagnósticos automáticos con los realizados por dos cardiólogos expertos de forma independiente, a partir de la lectura del registro realizado con el DMTE y de un ECG de 12 derivaciones.

Resultados:

participaron del estudio 114 beneficiarios del Plan Ibirapitá, 78 del sexo femenino. La edad fue de 72,5 ± 5,36 años (rango 53-87 años). La sensibilidad para el diagnóstico de FA del DMTE fue de 100%, con una especificidad de 96,6%. (VPP = 57,1% VPN = 100%) y una proporción de diagnóstico correcto de 96,8%. El DMTE catalogó como "sin clasificar" al 18,4% de los trazados. Considerando esto, la proporción de diagnóstico correcto disminuyó a 78,9%, sin presentar falsos negativos.

Conclusiones:

el cribado de FA con la utilización de un DMTE en una población de adultos mayores es factible y confiable. El hallazgo de un 18,4% de trazados con diagnóstico automático "sin clasificar" hace necesario contar con recursos humanos calificados para realizar la confirmación diagnóstica en esos registros.
ABSTRACT
Atrial fibrillation (AF) is a public health problem generating important morbidity and mortality mainly in > 65 years old population. Detection is key in the population at risk. This pilot study was designed to assess a mobile electronic technology device (METD) usefulness in AF screening.

Objective:

evaluate validity and reliability of a METD in AF identification. Secondary

objective:

to validate the process of collection, transmission, storage, method and interpretation of obtained data.

Method:

participants in 5 Ibirapitá Plan workshops (October 15-November 30, 2018) were invited to participate in the study. A KardiaMobile Alive Cor® METD was used, whose ECG recording was compared with a 12L ECG taken almost at the same time. Automatic METD report was compared with 2 independent cardiologists report considering the METD recording and the 12L ECG.

Results:

114 participants (78 female), mean age 72,5 y.o. (SD 5,36; range 53-87) were included. METD sensitivity for AF diagnosis was 100% with a 96.6% specificity (PPV=57,1% NPV=100%), and a 96.8% number of correct diagnosis. A 18.4% of recordings were catalogued as "unclassified" by the METD. Considering this, the proportion of correct diagnosis decreased to 78.9%; there were not false negatives.

Conclusions:

AF screening with a METD in an older population is feasible and reliable. The finding of 18.4% METD recordings as "unclassified" raises the needs for experts review during diagnosis confirmation.
RESUMO
A fibrilação atrial (FA) é um problema de saúde pública que gera mortalidade e morbidade, com maior impacto em pessoas com mais de 65 anos. Sua detecção é especialmente relevante para a população de risco. Este estudo piloto teve como objetivo avaliar a utilidade de um dispositivo móvel de tecnologia eletrônica (DMTE) para o monitoramento da FA. Objetivo principal avaliar a validade e confiabilidade de um DMTE para identificar FA. Objetivo secundário validar os processos de coleta, transmissão, armazenamento, processamento e interpretação dos dados obtidos.

Método:

participantes de 5 oficinas do Plano Ibirapitá (entre 15 de outubro e 30 de novembro de 2018) foram convidados a participar do estudo. Foi utilizado um DMTE (sensor AliveCor® KardiaMobile), cujo registro foi contrastado com o traçado do ECG de 12 derivações realizado quase simultaneamente. Os diagnósticos automatizados foram comparados com aqueles realizados por dois cardiologistas especialistas independentes, com base na leitura gráfica do DMTE e no ECG de 12 derivações.

Resultados:

participaram do estudo 114 beneficiários do Plano Ibirapitá, sendo 78 mulheres. A média de idade foi de 72,5 anos (DP 5,36; variação de 53-87 anos). A sensibilidade para o diagnóstico de FA no DMTE foi de 100% com especificidade de 96,6%. (VPP = 57,1% VPN = 100%) e proporção de diagnóstico correto de 96,8%. O DMTE definiu 18,4% dos registros como "não classificados". Portanto, a proporção de diagnósticos corretos diminuiu para 78,9% e não houve falsos negativos.

Conclusões:

a triagem para FA por meio de DMTE em uma população idosa é viável e confiável. A constatação de 18,4% dos registros com diagnóstico automático "não classificado" torna necessária a existência de recursos humanos qualificados para a realização da confirmação diagnóstica.
Assuntos
Palavras-chave

Texto completo: 1 Índice: LILACS Assunto principal: Fibrilação Atrial / Eletrocardiografia / Telemonitoramento Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies / Screening_studies Limite: Aged / Female / Humans / Male Idioma: Es Revista: Rev. urug. cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Índice: LILACS Assunto principal: Fibrilação Atrial / Eletrocardiografia / Telemonitoramento Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies / Screening_studies Limite: Aged / Female / Humans / Male Idioma: Es Revista: Rev. urug. cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2021 Tipo de documento: Article