Measurement of loosening torque of prosthetic components in Morse taper connection implants
J. res. dent
;
2(4): [335-342], jul.-ago2014.
Artigo
em Inglês
| LILACS
| ID: biblio-1363435
ABSTRACT
AIM:
The aim of this study was to verify two different types of interfaces in Morse taper connection implants, relating the mechanical resistance to a loosening torque that is required to separate or move the prosthetic abutment. MATERIAL ANDMETHODS:
Ten Morse taper connection implants and their respective abutments were divided into two groups (n = 5), DH (double hexagon) interface and OI (octagonal implant) interface. A standard insertion of implants was performed in a stainless steel base, where each abutment received a sequence of two consecutive tightening torques at a 10-minute interval, followed by a loosening torque, which was measured using a digital torque gauge. The Student's t-test with a 5% significance level was used for data analysis.RESULTS:
The results showed a statistically significant difference in the loosening torque values compared with DH torque values (p < 0.05). However, the difference between tightening and loosening torque values was not statistically significant in the OI group (p = 0.465).CONCLUSION:
The study concluded that there was a difference in the torque required to loosen the abutment screws between the DH group and the OI group, with the latter showing better results.
Texto completo:
DisponíveL
Índice:
LILACS (Américas)
Assunto principal:
Implantes Dentários
/
Projeto do Implante Dentário-Pivô
Limite:
Feminino
/
Humanos
/
Masculino
Idioma:
Inglês
Revista:
J. res. dent
Assunto da revista:
Odontologia
/
Sa£de Bucal
Ano de publicação:
2014
Tipo de documento:
Artigo
País de afiliação:
Brasil
Instituição/País de afiliação:
Faculty of Dentistry of Bauru/BR
/
Federal University of Rio Grande do Sul/BR
/
University of Southern Santa Catarina/BR
Similares
MEDLINE
...
LILACS
LIS