Your browser doesn't support javascript.
loading
Fractura de catéter implantable con reservorio por síndrome de pinch-off / Totally implantable venous access device fracture due to pinch-off syndrome
Norese, Mariano; Andersen, Gustavo F; Sinagra, Diego L.
  • Norese, Mariano; Clínica Bazterrica. Unidad de Cardiología Intervencionista y Terapéutica. Equipo de Cirugía Vascular. Buenos Aires. AR
  • Andersen, Gustavo F; Clínica Bazterrica. Unidad de Cardiología Intervencionista y Terapéutica. Equipo de Cirugía Vascular. Buenos Aires. AR
  • Sinagra, Diego L; Clínica Bazterrica. Unidad de Cardiología Intervencionista y Terapéutica. Equipo de Cirugía Vascular. Buenos Aires. AR
Rev. argent. cir ; 113(2): 258-262, jun. 2021. graf
Artigo em Espanhol | LILACS-Express | LILACS, BINACIS | ID: biblio-1365483
RESUMEN
RESUMEN La fractura y embolización de un catéter implantable con reservorio es una complicación infrecuente, pero potencialmente grave. El síndrome pinch-off fue descripto en los accesos venosos subclavios, cuando la sección del catéter se produce por la compresión intermitente entre la primera costilla y la clavícula. Informamos el caso de una paciente oncológica a quien se le colocó un catéter implantable con reservorio por acceso percutáneo ecoguiado y control radioscópico en la vena subclavia. Luego de varias sesiones de quimioterapia comenzó con mal funcionamiento del catéter; la radiografía de tórax evidenció la fractura completa del catéter a nivel del espacio costoclavicular con embolia de un fragmento a las cavidades cardíacas. Se realizó la extracción endovascular del catéter fracturado me diante lazo de nitinol por acceso percutáneo femoral, sin complicaciones. Se analizan la incidencia, los factores de riesgo y las medidas para prevenir el síndrome de pinch- off en los catéteres implantables con reservorio.
ABSTRACT
ABSTRACT Fracture and migration of totally implantable venous access devices is a rare but potentially serious complication. Pinch-off syndrome was described in subclavian venous accesses when the intermittent compression of the catheter between the first rib and the clavicle produces catheter fracture. The report the case of a patient with cancer who underwent implantation of a totally implantable venous access device through the subclavian vein under ultrasound and radioscopy guidance. After several cycles of chemotherapy, the patient started with malfunctioning of the device. The chest X-ray showed a complete fracture of the catheter at the level of the costoclavicular space with migration of a fragment of the catheter to the right cardiac chambers. The fractured catheter was percutaneously removed via the right femoral vein using nitinol gooseneck snare without complications. The incidence of the pinch-off syndrome, risk factors and prevention measures in totally implantable venous access devices are analyzed.


Texto completo: DisponíveL Índice: LILACS (Américas) Tipo de estudo: Fatores de risco Idioma: Espanhol Revista: Rev. argent. cir Assunto da revista: Cirurgia Geral Ano de publicação: 2021 Tipo de documento: Artigo País de afiliação: Argentina Instituição/País de afiliação: Clínica Bazterrica/AR