Development and validation of analytical methods by HPLC for quality control of avermectins and milbemycins in bulk samples
Dutra, Flávia Viana Avelar; Teixeira, Leila Suleimara; Pires, Bruna Carneiro; Florez, Diego Hernando Ângulo; Teixeira, Roseane Andrade; Borges, Keyller Bastos.
  • Dutra, Flávia Viana Avelar; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
  • Teixeira, Leila Suleimara; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
  • Pires, Bruna Carneiro; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
  • Florez, Diego Hernando Ângulo; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
  • Teixeira, Roseane Andrade; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
  • Borges, Keyller Bastos; Universidade Federal de São João del-Rei. Departamento de Ciências Naturais. São João del-Rei. BR
Braz. J. Pharm. Sci. (Online) ; 58: e19587, 2022. tab, graf
Artigo em Inglês | LILACS | ID: biblio-1384015
ABSTRACT
This work aims to develop analytical methods using high-performance liquid chromatography with a diode array detector (HPLC-DAD) for analysis and quantification of avermectins (AVMs) and milbemycins (MBMs) in bulk samples. First, the methods were optimized and some parameters such as temperature, flow rate, injection volume and mobile phase with different proportions of solvents were evaluated. The best chromatographic conditions were obtained using the Phenomenex® C18 (150 × 4.60 mm, 5 µm) column at a temperature of 20 °C, flow rate of 1.2 mL min-1, injection volume of 20 µL, and detection at 250 nm. Acetonitrile ultrapure water (87 13, v/v) was used as mobile phase for moxidectin and eprinomectin, and acetonitrile methanol ultrapure water (53 35 12, v/v/v) for abamectin and ivermectin. Under these conditions satisfactory results were obtained, with appropriate limits of detection and quantification, acceptable linearity, precision, accuracy, and robustness. These methods satisfy the need for analytical methods for the multi-determination of MBMs and the B1a and B1b forms of AVMs by HPLC-DAD, which can be considered simple, effective, innovative and should aid in the development of the fiel
Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Métodos de Análise Laboratorial e de Campo / Cromatografia Líquida de Alta Pressão / Estudo de Validação Idioma: Inglês Revista: Braz. J. Pharm. Sci. (Online) Assunto da revista: Farmacologia / Terapˆutica / Toxicologia Ano de publicação: 2022 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Federal de São João del-Rei/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Métodos de Análise Laboratorial e de Campo / Cromatografia Líquida de Alta Pressão / Estudo de Validação Idioma: Inglês Revista: Braz. J. Pharm. Sci. (Online) Assunto da revista: Farmacologia / Terapˆutica / Toxicologia Ano de publicação: 2022 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Federal de São João del-Rei/BR