Avaliação comparativa entre analisadores hematológicos Mindray BC-6800 e URIT 5500, com ênfase na capacidade de detecção de blastos em pacientes portadores de leucemia aguda / Comparative evaluation between Mindray BC-6800 and URIT 5500 hematology analyzers, with emphasis on the ability to detect blasts in patients with acute leukemia
Ribeiro, Monyque Barbosa; Brito, Camila dos Santos; Brito Junior, Lacy Cardoso de.
  • Ribeiro, Monyque Barbosa; Universidade Federal do Pará. Laboratório de Patologia Clínica Dr. Paulo C. Azevedo. Belém, Pará. BR
  • Brito, Camila dos Santos; Laboratório Beneficente de Belém. Belém, Pará. BR
  • Brito Junior, Lacy Cardoso de; Universidade Federal do Pará. Belém, Pará. BR
Rev. med. (São Paulo) ; 101(2): e-172549, mar.-abr. 2022.
Artigo em Inglês, Português | LILACS-Express | LILACS | ID: biblio-1390777
RESUMO
Comparar a capacidade dos analisadores hematológicos BC-6800 (Mindray) e URIT 5500 em sinalizar a presença de blastos em pacientes portadores de leucemia aguda. Foram analisadas 13 amostras de sangue periférico contendo blastos mielóide ou linfóide, provenientes de um hospital oncológico de Belém ­ Pará, previamente imunofenotipados por citometria de fluxo para verificar a capacidade dos equipamentos Mindray BC-6800 e URIT 5500 em sinalizar a presença dessas células no scatter leucocitário ou por emissão de ÀDJV. Para avaliação da existência de diferença estatística entre os resultados de hemácias, hemoglobina, leucócitos e plaquetas obtidos pelos equipamentos BC 6800 (Mindray) e URIT 5500 foi aplicado o teste não paramétrico ANOVA para análise de variância das amostras, o qual mostrou que não havia diferença estatística entre esses analitos. Não foi aplicado método estatístico para as contagens da diferencial leucocitária, pois o equipamento URIT 5500 não gerou dados numéricos para as amostras patológicas. Os dois equipamentos foram capazes de gerar ÀDJV e mudanças espacial do scatter leucocitário para amostras patológicas, contudo, o analisador BC 6800 (Mindray) foi o único a mudar a cor da população de blastos no scatter leucocitário. Os analisadores BC-6800 (Mindray) e URIT 5500 mostraram boa capacidade em sinalizar, através ÀDJV e do scatter leucocitário, para a presença de blastos mielóides ou linfóides em amostras patológicas. [au]
ABSTRACT
Compare the ability of the BC-600 (Mindray) and URIT 5500 hematological analyzers to signal the presence of blasts in patients with acute leukemia. Thirteen samples of peripheral blood containing myeloid or lymphoid blasts, from a cancer hospital in Belém - Pará, previously immunophenotyped by flow cytometry were analyzed to determine the capacity of the Mindray BC-6800 and URIT 5500 equipment in signaling the presence of these cells in the leukocyte scatter or by emitting flags. To assess the existence of statistical difference between the results of red blood cells, hemoglobin, leucocytes and platelets obtained by the BC 6800 (Mindray) and URIT 5500 equipments, the non-parametric ANOVA test was applied for analysis of variance of the samples, which showed that there was no statistical difference between these analytes. Statistical method was not applied for leukocyte differential counts, as the URIT 5500 equipment did not generate numerical data for the pathological samples. Both were able to generate flags and spatial changes from the leukocyte scatter to pathological samples, however the BC 6800 (Mindray) analyzer was the only one to change the color of the blast population in the leukocyte scatter BC-6800 (Mindray) and URIT 5500 analyzers showed good ability to signal, through flags and leukocyte scatter, for the presence of myeloid or lymphoid blasts in patholical samples. [au]