Influência do hipoclorito de sódio na resistência à fadiga cíclica em instrumentos rotatórios endodônticos de memória controlada de NiTi: uma avaliação experimental / Influence of sodium hypochlorite on cyclic fatigue resistance in niti controlled memory endodontics rotary instruments: an experimental evaluation
Rev. odontol. UNESP (Online)
; 51: e20220046, 2022. tab, ilus
Article
em Pt
| LILACS, BBO
| ID: biblio-1424236
Biblioteca responsável:
BR39.2
RESUMO
Introdução:
É importante saber se o hipoclorito de sódio (NaOCl) influencia a resistência à fadiga cíclica das limas de níquel-titânio (NiTi).Objetivo:
Avaliar a influência de NaOCl 2,5% na resistência à fadiga cíclica de dois sistemas de NiTi. Material emétodo:
40 instrumentos rotatórios - 20 TruNatomy® (TRU, Dentsply Sirona, Maillefer, Ballaigues, Suíça) e 20 Prodesing Logic2® (PDL2, Bassi, Belo Horizonte, Minas Gerais, Brasil) - foram aleatoriamente distribuídos em 4 grupos experimentais (n = 10) imersos em água destilada (H2O) e NaOCl 2,5% em temperatura a 37°C. Foram submetidos a testes de fadiga cíclica mensurando o número de ciclos para fratura (NCF) e análise dessas superfícies pós-teste em microscópio eletrônico por varredura. Para a análise estatística entre os grupos, foi aplicada a análise de variância (ANOVA), complementada com o pós-teste de Tukey.Resultado:
Houve diferença estatística em todos os grupos (P<0.05). Os instrumentos PDL2 obtiveram maior resistência à fratura nas condições em H2O e em NaOCl 2,5% comparados aos instrumentos TRU. Na análise de grupos de instrumentos nas soluções de NaOCl e H2O, foi observado que o NaOCl 2,5% diminuiu o NCF.Conclusão:
A resistência à fadiga cíclica dos instrumentos TRU e PDL2 diminuiu com NaOCl 2,5%. Os instrumentos PDL2 foram mais resistentes à fratura em relação aos instrumentos TRU.ABSTRACT
Introduction:
It is important to know whether sodium hypochlorite (NaOCl) influences the cyclic fatigue resistance of nickel-titanium (NiTi) files.Objective:
To evaluate the influence of NaOCl 2.5% on the cyclic fatigue resistance of two NiTi systems. Material andmethod:
Forty rotary instruments - 20 TruNatomy® (TRU, Dentsply Sirona, Maillefer, Ballaigues, Switzerland) and 20 Prodesign Logic2® (PDL2, Bassi, Belo Horizonte, Minas Gerais, Brazil) - were distributed randomly across four experimental groups (n=10) and submerged in distilled water (H2O) or 2.5% NaOCl at 37°C according to allocation. Cyclic fatigue testing was then performed, measuring the number of cycles to fracture (NCF), and post-test surfaces were analysed by scanning electron microscopy (SEM). StatisticalAnalysis:
Analysis of variance (ANOVA) was applied for between-group analysis, followed by Tukey's post-hoc test.Result:
A significant difference was observed in all groups (P<0.05). PDL2 instruments showed higher fracture resistance under H2O and 2.5% NaOCl conditions compared to TRU. Analysis of all instrument groups showed that exposure to 2.5% NaOCl decreased the NCF compared to H2O.Conclusion:
Cyclic fatigue resistance of the TRU and PDL2 instruments was decreased by exposure to 2.5% NaOCl. PDL2 instruments were more resistant to fracture than TRU instruments.Palavras-chave
Texto completo:
1
Índice:
LILACS
Assunto principal:
Hipoclorito de Sódio
/
Titânio
/
Microscopia Eletrônica de Varredura
/
Análise de Variância
/
Instrumentos Odontológicos
/
Níquel
Tipo de estudo:
Clinical_trials
Idioma:
Pt
Revista:
Rev. odontol. UNESP (Online)
Assunto da revista:
ODONTOLOGIA
Ano de publicação:
2022
Tipo de documento:
Article