Influence of diameter in the stress distribution of extra-short dental implants under axial and oblique load: a finite element analysis
Vargas-Moreno, Vanessa Felipe; Gomes, Rafael Soares; Ribeiro, Michele Costa de Oliveira; Freitas, Mariana Itaborai Moreira; Del Bel Cury, Altair Antoninha; Marcello-Machado, Raissa Micaella.
  • Vargas-Moreno, Vanessa Felipe; University of Campinas. Piracicaba Dental School. Department of Prosthodontics and Periodontology. Piracicaba. BR
  • Gomes, Rafael Soares; Faculty of Technology and Sciences. Department of Prosthodontics. Salvador. BR
  • Ribeiro, Michele Costa de Oliveira; University of Campinas. Piracicaba Dental School. Department of Prosthodontics and Periodontology. Piracicaba. BR
  • Freitas, Mariana Itaborai Moreira; University of Campinas. Piracicaba Dental School. Department of Prosthodontics and Periodontology. Piracicaba. BR
  • Del Bel Cury, Altair Antoninha; University of Campinas. Piracicaba Dental School. Department of Prosthodontics and Periodontology. Piracicaba. BR
  • Marcello-Machado, Raissa Micaella; University of Campinas. Piracicaba Dental School. Department of Prosthodontics and Periodontology. Piracicaba. BR
Braz. j. oral sci ; 22: e238152, Jan.-Dec. 2023. ilus
Artigo em Inglês | LILACS, BBO | ID: biblio-1437694
ABSTRACT

Aim:

This study evaluated the influence of a wide diameter on extra-short dental implant stress distribution as a retainer for single implant-supported crowns in the atrophic mandible posterior region under axial and oblique load.

Methods:

Four 3D digital casts of an atrophic mandible, with a single implant-retained crown with a 31 crown-to-implant ratio, were created for finite element analysis. The implant diameter used was either 4 mm (regular) or 6 mm (wide), both with 5 mm length. A 200 N axial or 30º oblique load was applied to the mandibular right first molar occlusal surface. The equivalent von Mises stress was recorded for the abutment and implant, minimum principal stress, and maximum shear stress for cortical and cancellous bone.

Results:

Oblique load increased the stress in all components when compared to axial load. Wide diameter implants showed a decrease of von Mises stress around 40% in both load directions at the implant, and an increase of at least 3.6% at the abutment. Wide diameter implants exhibited better results for cancellous bone in both angulations. However, in the cortical bone, the minimum principal stress was at least 66% greater for wide than regular diameter implants, and the maximum shear stress was more than 100% greater.

Conclusion:

Extra-short dental implants with wide diameter result in better biomechanical behavior for the implant, but the implications of a potential risk of overloading the cortical bone and bone loss over time, mainly under oblique load, should be investigated

Assuntos


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Implantes Dentários / Arcada Parcialmente Edêntula / Prótese Dentária Fixada por Implante / Análise de Elementos Finitos Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2023 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculty of Technology and Sciences/BR / University of Campinas/BR

Similares

MEDLINE

...
LILACS

LIS


Texto completo: DisponíveL Índice: LILACS (Américas) Assunto principal: Implantes Dentários / Arcada Parcialmente Edêntula / Prótese Dentária Fixada por Implante / Análise de Elementos Finitos Idioma: Inglês Revista: Braz. j. oral sci Assunto da revista: Odontologia Ano de publicação: 2023 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculty of Technology and Sciences/BR / University of Campinas/BR