Estimación de Distensibilidad Dinámica Toraco-pulmonar en Modelos Neonatales de Ventilación Mecánica de Alta Frecuencia / Estimation of dynamic thoracopulmonary compliance in neonatal high-frequency mechanical ventilation models
Adasme Jeria, Rodrigo; González Oyarzo, María Constanza; Manzanares López, Javier; Ormeño Cid, Stephania; Zelada Pellegrin, Karen; Figueroa González, Paola; Rodrigo Adasme Jeria.
  • Adasme Jeria, Rodrigo; Universidad Andrés Bello. Faculty of Rehabilitation Sciences. Exercise and Rehabilitation Sciences Institute, School of Physical Therapy. Santiago. CL
  • González Oyarzo, María Constanza; Universidad Andrés Bello. Faculty of Rehabilitation Sciences. Exercise and Rehabilitation Sciences Institute, School of Physical Therapy. Santiago. CL
  • Manzanares López, Javier; Universidad Andrés Bello. Faculty of Rehabilitation Sciences. Exercise and Rehabilitation Sciences Institute, School of Physical Therapy. Santiago. CL
  • Ormeño Cid, Stephania; Universidad Andrés Bello. Faculty of Rehabilitation Sciences. Exercise and Rehabilitation Sciences Institute, School of Physical Therapy. Santiago. CL
  • Zelada Pellegrin, Karen; Universidad Andrés Bello. Faculty of Rehabilitation Sciences. Exercise and Rehabilitation Sciences Institute, School of Physical Therapy. Santiago. CL
  • Figueroa González, Paola; Universidad Andrés Bello. Facultad de Ciencias de la Rehabilitación. Escuela de Kinesiología. Santiago. CL
Kinesiologia ; 42(4): 275-284, 20231215.
Artigo em Espanhol, Inglês | LILACS-Express | LILACS | ID: biblio-1552535
RESUMEN
Introducción. La ventilación de alta frecuencia (VAF) es utilizada en pacientes graves neonatales con un uso cercano al 10% del total de usuarios en ventilación mecánica (VM). Actualmente estos equipos miden volumen corriente de manera precisa, continua y rutinaria, sin embargo no hay estimaciones del comportamiento mecánico del sistema respiratorio del usuario, como lo es la distensibilidad toracopulmonar, durante el ciclo ventilatorio que sean reportadas por las pantallas de los equipos. Objetivo. Estimar distensibilidad dinámica toracopulmonar en modelos neonatales de VAF. Métodos. Estudio cuantitativo, observacional, descriptivo, y "wench work", donde se midió distensibilidad en VM convencional y volumen corriente (Vt) en VAF con equipo Acutronics Fabian® por 4 evaluadores independientes, en 5 tipos de pulmón de ensayo y bajo diferentes escenarios de parámetros de VAF con ajustes de presión media de la vía aérea (PMVA), amplitud, frecuencia respiratoria, tiempo inspiratorio, volumen garantizado, y tipo de circuito. Mediante suavización de regresiones locales por estimación mínima cuadrática (LOWESS) y análisis de regresión multivariada se obtuvieron los valores asociados a distensibilidad, con los que se construyeron ecuaciones de estimación de distensibilidad en VAF. Resultados. Se realizaron en total 46080 mediciones, con correlación intra e interevaluador > 0.99. La distensibilidad mediana (percentil 25; 75) de los 5 modelos pulmonares fue de 0.455 (0.3; 0.98). Se asociaron a distensibilidad, mediante modelos de regresión lineal múltiple de manera significativa, todas las variables evaluadas, salvo PMVA. Se establecieron asociaciones multivariantes crudas (R2=.311), de distensibilidad predicha por LOWESS (R2=.744) y distensibilidad predicha y variables independientes predichas por regresión lineal simple (R2=.973). Conclusiones. La distensibilidad en VAF esta determinada en pulmones de ensayos por los parámetros programados de tipo de circuito, uso de volumen garantizado, tiempo inspiratorio, frecuencia respiratoria y amplitud, además del Vt medido. Se reporta ecuación explicativa de distensibilidad en VAF.
ABSTRACT
Background.

Background:

High frequency mechanical ventilation (HFV) is used in severe neonatal subjects nearly 10% of total mechanically ventilated (MV) users. Currently, this MV's mode allow to measure tidal volumen in an accurately, continuous and routinarie way in HFV, however there are non estimation to assess mechanical behavior of respiratory system during ventilatory cycle, like thoraco-pulmonary compliance, who be report in the equipment display. Objetive. To estimate thoraco-pulmonary compliance in artificial neonatal lung models in HFV. Methods. Quantitative, observational, descriptive, and wench work study, where distensibility was measured in conventional MV and tidal volume (TV) in HFV with Acutronics Fabian® equipment by 4 independent evaluators, in 5 types of test lung and under different scenarios of HFV parameters with adjustments of mean airway pressure (MAP), amplitude, respiratory rate, inspiratory time, guaranteed volume, and type of circuit. By smoothing local regressions by least quadratic estimation (LOWESS) and multivariate regression analysis, the values associated with distensibility were obtained, with this measures, equations for estimating compliance in VAF were constructed. Results. A total of 46080 measurements were made, with intra and inter-evaluator correlation coefficent > 0.99. The median compliance (25th percentile; 75) of the 5 lung models was 0.455 (0.3; 0.98). All variables evaluated, except MAP, were associated with compliance, by means of multiple linear regression models. Crude multivariate associations (R2 = .311), predicted compliance by LOWESS (R2 = .744) and predicted compliance and independent variables predicted by simple linear regression (R2 = .973) were established to estimate thoraco-pulmonary compliance. Conclusions. Compliance in VAF is determined in test lungs by the programmed parameters of type of circuit, use of guaranteed volume, inspiratory time, respiratory frequency and amplitude, in addition to the measured Vt. An explanatory equation for distensibility in VAF is reported.

Texto completo: DisponíveL Índice: LILACS (Américas) Idioma: Inglês / Espanhol Revista: Kinesiologia Assunto da revista: Medicina Ano de publicação: 2023 Tipo de documento: Artigo País de afiliação: Chile Instituição/País de afiliação: Universidad Andrés Bello/CL