Your browser doesn't support javascript.
loading
Efeito da soldagem híbrida (plasma + maçarico) na suavização da assinatura torque-ângulo em supraestruturas de COCR sobre minipilares de zircônia ­ parte 1 / Effect of hybrid soldering (plasma + gas torch) technique on suavization of torque-angle signatures at Co-Cr frameworks over zirconia mini-abutmets ­ part I
Campos, Adriana Lima; Bonachela, Wellington Cardoso; Rossetti, Paulo Henrique Orlato; Brito Jr, Rui Barbosa de; Bonachela, Clara Fassoni; Marques, Vinicius Rizzo.
  • Campos, Adriana Lima; s.af
  • Bonachela, Wellington Cardoso; USP. Faculdade de Odontologia de Bauru. Depto. de Prótese e Periodontia. Bauru. BR
  • Rossetti, Paulo Henrique Orlato; s.af
  • Brito Jr, Rui Barbosa de; Faculdade São Leopoldo Mandic. Depto. de Prótese. Campinas. BR
  • Bonachela, Clara Fassoni; s.af
  • Marques, Vinicius Rizzo; s.af
ImplantNewsPerio ; 1(4): 712-721, mai.-jun. 2016. ilus, tab
Artigo em Português | LILACS, BBO | ID: biblio-847034
RESUMO
Objetivo: avaliar as desadaptações em supraestruturas de implantes e a curva de assinatura torque-ângulo dos parafusos protéticos nas condições monobloco e pós-soldagem. Material e métodos: uma base retangular de aço inox recebeu três implantes de hexágono externo (4,1 mm x 10 mm), onde foram parafusados minipilares de zircônia. Após o enceramento e inclusão, as supraestruturas foram fundidas em monobloco com uma liga de cobalto-cromo, e os cilindros foram numerados sequencialmente (1, 2 e 3). A leitura na interface supraestrutura/pilar foi realizada com um microscópio comparador (precisão de 1 µm), três vezes em cada cilindro. A tensão de torque nos parafusos foi medida com um torquímetro eletrônico odontológico (OsseoCare, Nobel Biocare). Depois, as supraestruturas foram seccionadas e soldadas a plasma (pontos de estabilização) e maçarico (preenchimento da área restante). Novamente, as desadaptações e assinaturas dos parafusos foram avaliadas pela mesma metodologia. Resultados: as médias de desadaptações foram maiores nas supraestruturas em monobloco (C1=3,5 µm; C2=0 µm; C3=31,2 µm) do que nas supraestruturas pós-soldagem (C1=3,0 µm; C2=2,9 µm; C3=18,1 µm). Dentro de cada condição, o teste de Kruskal-Wallis mostrou diferença estatisticamente significativa apenas para o cilindro 3 (monobloco: p < 0,00003 / pós-soldagem: p=0,008) em relação aos cilindros 1 e 2. Uma diferença estatisticamente significante foi encontrada apenas no C3, comparando as condições monobloco e pós-soldagem (teste t pareado, p=0,03). A assinatura dos parafusos se mostrou melhor na condição pós-soldagem. Conclusão: a fundição monobloco gera instabilidade na assinatura dos parafusos. A desadaptação nos parafusos protéticos de supraestruturas sobre minipilares de zircônia melhora após secção e soldagem híbrida.
ABSTRACT
Objective: to evaluate the misfi t at implant frameworks and the torque-angle signature curves at the prosthetic screws at one-piece and after soldering conditions. Material and methods: a stainless steel rectangular base received 3 external hex implants (4.1 x 10 mm) where zirconia mini-abutments were fastened. After wax-up and investing, the frameworks were one-piece cast with a Co-Cr alloy, and the prosthetic cylinders sequentially identified (1, 2, and 3). The misfit at the framework/abutment interface was measured with a microscope (reading error 1 µm) 3 times for each cylinder. The screw tension was investigated with an electronic torqued device (Osseocare, Nobel Biocare). After, the frameworks were sectioned and soldered with plasma (stabilization points) and gas torch (filling of the remaining areas). Again, the misfit and torque-angle signatures were measured as described. Results: mean misfit values were greater for one-piece castings (C1=3.5 µm; C2=0 µm; C3=31.2 µm) than after soldering (C1=3.0 µm; C2=2.9 µm; C3=18.1 µm). Within each condition, the Kruskal-Wallis test demonstrated a statistically significant difference only for C3 (one-piece casting: p < 0.00003 / after soldering: p=0.008) compared to C1 and C2. Between each condition, a significant difference was seen only for C3 in the one-piece and after soldering conditions (paired t test, p=0.03). The torque-angle signatures demonstrated a better behavior after soldering. Conclusion: one-piece castings provide instability at torque-angle signatures. Thus, the seating of the prosthetic abutments over the zirconia mini-abutments improves after sectioning and hybrid soldering.
Assuntos

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Soldagem em Odontologia / Zircônio / Implantes Dentários / Ligas Metalo-Cerâmicas / Torque Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculdade São Leopoldo Mandic/BR / USP/BR

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Soldagem em Odontologia / Zircônio / Implantes Dentários / Ligas Metalo-Cerâmicas / Torque Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Faculdade São Leopoldo Mandic/BR / USP/BR