Desajuste vertical entre implante de hexágono interno e pilares Ucla totalmente calcináveis e com base de CoCr após a fundição pelo processo convencional / Vertical misfit between internal hexagon implants and Ucla burnout and CoCr prefabricated base abutments after conventional casting procedures
Dayube, Ulisses Ribeiro Campos; Furtado, Thayane Silveira Mata; Leal, Thales de Assis Brasil; Cunha, Rodrigo Moreira; Isabel, Carlos Alberto Camargo; Shibli, Jamil Awad.
  • Dayube, Ulisses Ribeiro Campos; s.af
  • Furtado, Thayane Silveira Mata; s.af
  • Leal, Thales de Assis Brasil; s.af
  • Cunha, Rodrigo Moreira; s.af
  • Isabel, Carlos Alberto Camargo; s.af
  • Shibli, Jamil Awad; Universidade Guarulhos (UNG). Área de Implantodontia e Periodontia. Guarulhos. BR
ImplantNewsPerio ; 2(2): 227-233, mar.-abr. 2017. ilus, tab
Artigo em Português | LILACS, BBO | ID: biblio-847144
RESUMO

Objetivo:

avaliar, por meio de um microdurômetro, o grau de desajuste na adaptação de pilares Ucla em implantes. Material e

métodos:

foram utilizados dez implantes de hexágono interno de plataforma regular, e seus respectivos componentes protéticos do tipo Ucla com antirrotacional, como seguem dez pilares Ucla totalmente calcináveis e dez pilares Ucla com base de CoCr. Todos os pilares Ucla foram fundidos pelo mesmo laboratório de prótese e pelo mesmo processo laboratorial. Cada implante recebeu apenas dois pilares, que foram anexados por meio de um parafuso de titânio quadrado utilizando-se torque de 30 Ncm com torquímetro manual. Cada conjunto implante/pilar recebeu quatro marcações aleatórias em torno da circunferência, na qual foram feitas as medições do espaço existente entre as peças utilizando uma lupa estereoscópica de 100 vezes de aumento.

Resultados:

os resultados obtidos mostraram diferenças significativas entre os grupos submetidos ao teste Mann-Whitney (p < 0,05).

Conclusão:

o pilar Ucla com base de CoCr apresentou menor desajuste marginal vertical, quando comparado ao pilar Ucla totalmente calcinável, após o processo de fundição nos implantes de hexágono interno.
ABSTRACT

Objective:

to evaluate, using a microhardness device, the misfit between Ucla abutments and internal hex implants. Material and

methods:

10 internal hex implants received 10 Ucla burnout and 10 Ucla abutments with CoCr collar. All the abutments were cast in the same laboratory and under the same protocol. A square-head titanium screw was tightened to a 30 Ncm using a manual wrench. Each abutment/implant set receive four random markings around its perimeter. A stereoscopic lens was used to investigate the misfi t at 100 magnification.

Results:

there was a statistically significant difference between tested abutments (Mann-Whitney p < 0.05).

Conclusion:

the Ucla abutment with the CoCr collar presented less vertica misfit values after casting for internal hex dental implants.
Assuntos

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Interpretação Estatística de Dados / Técnica de Fundição Odontológica / Planejamento de Prótese Dentária / Prótese Dentária Fixada por Implante / Revestimento para Fundição Odontológica Tipo de estudo: Ensaio Clínico Controlado / Estudos de avaliação Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Guarulhos (UNG)/BR

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Interpretação Estatística de Dados / Técnica de Fundição Odontológica / Planejamento de Prótese Dentária / Prótese Dentária Fixada por Implante / Revestimento para Fundição Odontológica Tipo de estudo: Ensaio Clínico Controlado / Estudos de avaliação Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Guarulhos (UNG)/BR