Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Iranian Journal of Radiation Research. 2010; 8 (3): 195-199
em Inglês | IMEMR | ID: emr-109518
ABSTRACT
Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran [SSDL] has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. The present work has described traceability of SSDL radiation measurement standards to relevant international standards, and calibration of therapy level ionization chambers in terms of air kerma and absorbed dose to water against [60]Co gamma radiation, as well as uncertainty evaluation of calibration coefficients. The expanded uncertainties in the determination of air kerma and absorbed dose to water are estimated to be 2% and 2.3% at approximately 95% confidence level, respectively. In order to maintain the requirement of +/- 5% accuracy in the dose delivery, the combined standard uncertainty of the other factors in the dose delivery; i.e., dose measurement set-up, dose calculations, treatment planning, patient set-up, etc, should be less than 2.3%
Assuntos
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Radiometria / Calibragem / Incerteza Idioma: Inglês Revista: Iran. J. Radiat. Res. Ano de publicação: 2010

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Radiometria / Calibragem / Incerteza Idioma: Inglês Revista: Iran. J. Radiat. Res. Ano de publicação: 2010