Live/real time three-dimensional trans esophageal echocardiographic findings in amplatzer ASD closure devices in adults
International Cardiovascular Research Journal. 2012; 6 (3): 97-100
em Inglês
| IMEMR
| ID: emr-153991
ABSTRACT
Six female patients aged from 19 to 73 years, with ostium secundum atrial septal defect underwent closure procedure with Amplatzer septal occluder device. Three-dimensional Echocardiography [3D-TEE] was done during the procedure or one day after the procedure. 3D-TEE provides incremental value over Two- dimensional trans-esophageal echocardiography in measuring Amplatzer septal occluder disc sizes and correlates well with manufacture device size. 3D-TEE will surely prove to increase the technical efficiency and it will become an important tool for the interventionists for periprocedural evaluation of device closures