Your browser doesn't support javascript.
loading
Case report of deep vein thrombosis caused by artificial urinary sphincter reservoir compressing right external iliac vein
Urology Annals. 2015; 7 (1): 112-114
em Inglês | IMEMR | ID: emr-154922
ABSTRACT
Artificial urinary sphincters [AUSs] are commonly used after radical prostatectomy for those who are incontinent of urine. However, they are associated with complications, the most common being reservoir uprising or migration. We present a unique case of occlusive external iliac and femoral vein obstruction by the AUS reservoir causing thrombosis. Deflation of the reservoir and anticoagulation has, thus far, not been successful at decreasing thrombus burden. We present this case as a rare, but significant surgical complication; explore the risk factors that may have contributed, and other potential endovascular therapies to address this previously unreported AUS complication
Assuntos
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Tomografia Computadorizada por Raios X / Esfíncter Urinário Artificial / Veia Ilíaca Tipo de estudo: Relato de Casos Limite: Humanos / Masculino Idioma: Inglês Revista: Urol. Ann. Ano de publicação: 2015

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Tomografia Computadorizada por Raios X / Esfíncter Urinário Artificial / Veia Ilíaca Tipo de estudo: Relato de Casos Limite: Humanos / Masculino Idioma: Inglês Revista: Urol. Ann. Ano de publicação: 2015