Your browser doesn't support javascript.
loading
Comparison of Perkins, Tono-Pen and Schiotz tonometers in paediatric patients under general anaesthesia
EMHJ-Eastern Mediterranean Health Journal. 2008; 14 (6): 1365-1371
em Inglês | IMEMR | ID: emr-157280
ABSTRACT
This cross-sectional study aimed to investigate if the Tono-Pen, Schiotz and Perkins tonometers could be used interchangeably in general practice for measuring elevated intraocular pressure. A total of 74 eyes of 37 paediatric patients under general anaesthesia were checked with all 3 tonometers. All of the tonometers gave significantly different measurements from each other. However, with a mean difference of 1.4 mmHg and 95% limits of agreement of -5.7 to +8.6, the greatest agreement was between the Perkins and Tono-Pen tonometers. The Perkins tonometer is a hand-held variant of the Goldmann tonometer [the gold standard for intraocular pressure measures]. Therefore the Tono-Pen with its ease of use and safety could be a reliable device for use in general practice
Assuntos
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Glaucoma / Estudos Transversais / Fatores de Risco / Diagnóstico Precoce / Pressão Intraocular / Anestesia Geral Tipo de estudo: Estudo de prevalência / Estudo de rastreamento Limite: Feminino / Humanos / Masculino Idioma: Inglês Revista: East Mediterr Health J. Ano de publicação: 2008

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Glaucoma / Estudos Transversais / Fatores de Risco / Diagnóstico Precoce / Pressão Intraocular / Anestesia Geral Tipo de estudo: Estudo de prevalência / Estudo de rastreamento Limite: Feminino / Humanos / Masculino Idioma: Inglês Revista: East Mediterr Health J. Ano de publicação: 2008