Falsely increased bispectral index values by convective air warming system during kidney transplantation
Pakistan Journal of Medical Sciences. 2016; 32 (3): 793-795
em Inglês | IMEMR | ID: emr-182988
ABSTRACT
Bispectral index [BIS] is a reliable parameter for measuring depth of hypnotic level during anesthesia. Convective air warming system is an effective equipment to maintain normothermia during operation. We report falsely elevated BIS value due to convective air warming system while undergoing kidney transplantation
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Idioma: Inglês Revista: Pak. J. Med. Sci. Ano de publicação: 2016