Transcutaneous bilirubinometery: its role in the assessment of neonatal jaundice with comparable precision with serum bilirubin
Alexandria Journal of Pediatrics. 2004; 18 (1): 327-333
em Inglês | IMEMR | ID: emr-201171
ABSTRACT
Transcutaneous bilirubinometry [TCB] provides a noninvasive, cost effective screening method for significant neonatal jaundice. The aim of this study is to comment on the usefulness of the jaundice meter as a non invasive screening device for hyperbilirubinemia in neonatal intensive care units. The study included 40 neonates suspected on clinical grounds to have neonatal jaundice, admitted to the unit for either prematurity or other clinical conditions. Our results showed a correlation coefficient between total serum bilirubin and the mean of transcutaneous bilirubinometry of the forehead and sternum at 24 and 48 hours of appearance of jaundice to be 0.629 and 0.823 respectively with a significant P value of [0.07. Also, the correlation coefficient between assayed total serum bilirubin level and expectant serum bilirubin level at 24 and 48 hrs. of appearance of jaundice was 0.599 and 0.863 respectively with a significant P value of <0.01
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Idioma: Inglês Revista: Alex. J. Pediatr. Ano de publicação: 2004