Medical Journal of Cairo University [The]. 1993; 61 (4): 783-787
em Inglês
| IMEMR
| ID: emr-29205
ABSTRACT
Readings from an Omron HEM 815F portable blood pressure monitor and a Dinamap 8100 are compared to those obtained from a mercury column sphygmomanometer in this study. Results showed a bias of -3.6 and -2.8 mmHg for systolic and diastolic readings respectively with the Omron HEM 815F, compared to Dinamap 8100 which showed bias of +11.9 and -3.2 mmHg. The limits of agreement for the systolic readings were +/- 23.8 mmHg from the bias [mean difference] with the Omron HEM 815F, compared to +/- 21.8 mmHg with the Dinamap 8100. The Omron HEM 815F is a small monitoring device that produces readings of comparable accuracy to those produced by the Dinamap 8100
Buscar no Google
Índice:
IMEMR (Mediterrâneo Oriental)
Assunto principal:
Oscilometria
/
Análise de Regressão
/
Estudo de Avaliação
Idioma:
Inglês
Revista:
Med. J. Cairo Univ.
Ano de publicação:
1993
Similares
MEDLINE
...
LILACS
LIS