Your browser doesn't support javascript.
loading
Electron beam therapy shielding material and clinical applications
Journal of the Medical Research Institute-Alexandria University. 1996; 17 (3): 88-95
em Inglês | IMEMR | ID: emr-41293
ABSTRACT
The relative percent dose reduction of different electron energies [5, 9 and 13 Mev] were measured using Mevatron 74 Linear accelerator and different shielding materials [lead, aluminium, copper and stainless steel]. A step wedge of each was used during this wrk, the steps thickness ranged between 2 mm and 20 mm. It was found that 95% attenuation for energies [5, 9 and 13 Mev] were obtained by 2, 4 and 6 mm thickness of lead respectively compared to the attenuation obtained by other materials of the same tickness
Assuntos
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Radioterapia / Cobre / Alumínio / Dosimetria Fotográfica / Chumbo Idioma: Inglês Revista: J. Med. Res. Inst.-Alex. Univ. Ano de publicação: 1996

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Radioterapia / Cobre / Alumínio / Dosimetria Fotográfica / Chumbo Idioma: Inglês Revista: J. Med. Res. Inst.-Alex. Univ. Ano de publicação: 1996