Your browser doesn't support javascript.
loading
Efficacy of esophageal detector device in verification of endotracheal tube placement
PAFMJ-Pakistan Armed Forces Medical Journal. 2006; 56 (1): 3-6
em Inglês | IMEMR | ID: emr-79875
ABSTRACT
To assess accuracy of esophageal detector device [EED] for detection of endotracheal tube placement and to compare its performance with that of capnography. A prospective study in which 400 patients were divided into two groups. In group-I the patients were intubated as per routine, then the EDD and capnograph were again used to check both the tubes. The result showed 100% sensitivity, 100% specificity and 100% predictive value for EDD. The EDD was found to be very effective device in differentiating esophageal from tracheal intubation
Assuntos
Buscar no Google
Índice: IMEMR (Mediterrâneo Oriental) Assunto principal: Estudos Prospectivos / Capnografia / Intubação / Intubação Intratraqueal Limite: Feminino / Humanos / Masculino Idioma: Inglês Revista: Pak. Armed Forces Med. J. Ano de publicação: 2006