The use of biosensors in pacemaker technology: experience with a ventricular rate responsive pacemaker
P. R. health sci. j ; 10(1): 15-8, Apr. 1991. tab
Artigo em Inglês | LILACS | ID: lil-100902
RESUMO
Ten elderly patientes each had a ventricular rate responsive pulse generator (Activitrax) placed in then to help correct severe conduction abnormalities; none had suspected ischemic heart disease. The pulse generator was programmed to a maximal pacing rate of 125 ppm, a medium activity threshold, and a rate response of 6. Six weeks after implantation of the pulse generator, the patients were evaluated before exercing and again when the pacing rate reached 125ppm. The evaluation protocol included an M-mode echocardiogram from which the following measurements were taken; the left ventricular end-diastolic volume (EDV), the end-systolic volume (ESV), the ejection fraction (EF), and the peak systolic pressure/end-systolic volume (PSP/ESV). The numerical values were recorded, clculated, and compared statically with the following

results:

the EDV increased from 91 ñ 10 to 125 ñ 20 cc (p < .05); the ESV decreased from 64 ñ 10 to 24 ñ 6cc (p < ..005); the EF increased from 41 ñ 5 to 61 ñ 10% (p < .05); and the PSP/ESV ratio increased from 1.70 ñ 1 to 4.10 ñ 2mm Hg/cc (p = 10). Also, during the maximal pacing rate, the septum of all patients showed paradoxical septal motion. All patients in our study have been asymptomatic and have shown an increase in their exercise capacity. We conclude that during exercise the left ventricular function ins influenced more by heart rate than by AV synchrony, as indicated by an elevated EDV in most patients
Assuntos
Buscar no Google
Índice: LILACS (Américas) Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Técnicas Biossensoriais / Frequência Cardíaca / Ventrículos do Coração Limite: Idoso / Humanos Idioma: Inglês Revista: P. R. health sci. j Assunto da revista: Medicina Ano de publicação: 1991 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Buscar no Google
Índice: LILACS (Américas) Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Técnicas Biossensoriais / Frequência Cardíaca / Ventrículos do Coração Limite: Idoso / Humanos Idioma: Inglês Revista: P. R. health sci. j Assunto da revista: Medicina Ano de publicação: 1991 Tipo de documento: Artigo