Experimental determination of bone cortex holding power of orthopedic screw
Bolliger Neto, Raul; Rossi, João D. M. B. Alvarenga; Leivas, Tomaz Puga.
Afiliação
  • Bolliger Neto, Raul; Universidade de Säo Paulo. Faculadade de Medicina. Instituto de Ortopedia e Traumatologia.
  • Rossi, João D. M. B. Alvarenga; Universidade de Säo Paulo. Faculadade de Medicina. Instituto de Ortopedia e Traumatologia.
  • Leivas, Tomaz Puga; Universidade de Säo Paulo. Faculdade de Medicina. Instituto de Ortopedia e Traumatologia.
Rev. Hosp. Clin. Fac. Med. Univ. Säo Paulo ; 54(6): 181-6, Nov.-Dec. 1999. ilus, tab, graf
Article em En | LILACS | ID: lil-260494
Biblioteca responsável: BR1.1
ABSTRACT
Cylindrical specimens of bone measuring 15 mm in diameter were obtained from the lateral cortical layer of 10 pairs of femurs and tibias. A central hole 3.2 mm in diameter was drilled in each specimen. The hole was tapped, and a 4.5 mm cortical bone screw was inserted from the outer surface. The montage was submitted to push-out testing up to a complete strip of the bone threads. The cortical thickness and rupture load were measured, and the shear stress was calculated. The results were grouped according to the bone segment from which the specimen was obtained. The results showed that bone cortex screw holding power is dependent on the bone site. Additionally, the diaphyseal cortical bone tissue is both quantitatively and qualitatively more resistant to screw extraction than the metaphyseal tissue.
Assuntos
Texto completo: 1 Índice: LILACS Assunto principal: Tíbia / Parafusos Ósseos / Fêmur Idioma: En Revista: Rev. Hosp. Clin. Fac. Med. Univ. Säo Paulo Assunto da revista: MEDICINA Ano de publicação: 1999 Tipo de documento: Article
Texto completo: 1 Índice: LILACS Assunto principal: Tíbia / Parafusos Ósseos / Fêmur Idioma: En Revista: Rev. Hosp. Clin. Fac. Med. Univ. Säo Paulo Assunto da revista: MEDICINA Ano de publicação: 1999 Tipo de documento: Article