Análise química e estrutural de componente femoral de prótese total do quadril removido após fratura por fadiga: relato de caso / Chemical and structural analysis of total hip prosthesis femoral component retrieved after fatigue fracture: a case report
Rev. bras. ortop
;
37(6): 256-258, jun. 2002. ilus, tab
Artigo
em Português
| LILACS
| ID: lil-322141
ABSTRACT
The authors present one case of stem fatigue fracture ofa cemented total hip prosthesis, which took place 43 months after its implantation. There was good cement thickness, adequate stem positioning, and absence of radiographic signs of loosening at the revision. The retrieved implant and an unused stem of the same type (control) were analyzed by macrographic inspection, metallographic and chemical analysis. The implants were found to have improper attributes of manufacturing and structural nature